| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 8171590 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 5 Pages |
Abstract
A new method for improving REM counter performance in Pulsed Neutron Fields (PNFs) has been developed. This method uses an analysis of the build-up of space charge in the counter to compensate for an underestimation of Ambient Dose Equivalent (H*(10)) in intense pulsed fields. It was applied to three sets of experimental data acquired using the LUPIN-II REM counter device, which is designed for use in PNFs. The data was acquired using the cyclotron at Helmholtz-Zentrum Berlin für Materialien und Energie GmbH (HZB), at the HiRadMat facility at CERN and at the 'Elettra Sincrotrone Trieste' (ELETTRA), Italy. A comparison of the data with and without this compensation method is used to highlight its effectiveness. The LUPIN-II performance, which has already been shown to be able to cope with fields of up to hundreds of nSv/burst, is improved by at least one order of magnitude, with further potential for improvement.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
C. Cassell, M. Ferrarini, A. Rosenfeld, M. Caresana,