Article ID Journal Published Year Pages File Type
8171607 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2015 7 Pages PDF
Abstract
Monoenergetic neutron response functions were measured for an EJ-299-33A plastic scintillator. The 7-MV Van de Graaff accelerator at the University of Kentucky Accelerator Laboratory was used to produce proton and deuteron beams for reactions with gaseous tritium and deuterium targets, yielding monoenergetic neutrons by means of the 3H(p,n)3He, 2H(d,n)3He, and 3H(d,n)4He reactions. The neutron energy was selected by tuning the charged-particle's energy and using the angular dependence of the neutron emission. The resulting response functions were measured for 0.1-MeV steps in neutron energy from 0.1 MeV to 8.2 MeV and from 12.2 MeV to 20.2 MeV. Experimental data were processed using a procedure for digital pulse-shape discrimination, which allowed characterization of the response functions of the plastic scintillator to neutrons only. The response functions are intended for use in neutron spectrum unfolding methods.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,