| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 8172256 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 7 Pages | 
Abstract
												The beam characterization studies in terms of emittance and proton fraction have been carried out in the high current Electron Cyclotron Resonance (ECR) proton ion source developed for Low Energy High Intensity Proton Accelerator (LEHIPA). The beam emittance was measured using two slit emittance measurement units (EMU). The emittance was measured at three locations (1) after beam extraction at ion source end, (2) after focusing the beam using solenoid magnet and (3) after focusing and separating H+ using solenoid magnet and analyzing magnet. The beam emittance measured in all three cases was found to be less than 0.2Ï mm-mrad (rms-normalized). The proton fraction in the beam measured using analyzing magnet was found to be more than 90%. The variations of beam emittance and proton fraction have been studied as a function of microwave power and neutral gas pressure.
											Keywords
												
											Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Instrumentation
												
											Authors
												P. Roychowdhury, H. Kewlani, L. Mishra, S. Gharat, R.K. Rajawat, 
											