Article ID Journal Published Year Pages File Type
8172554 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2015 7 Pages PDF
Abstract
Where energy resolution is paramount, High Purity Germanium (HPGe) detectors continue to provide the optimum solution for gamma-ray detection and spectroscopy. Conventional large-volume HPGe detectors are typically limited to count rates on the order of ten thousand counts per second, however, limiting their effectiveness for high count rate applications. To address this limitation, we have developed a novel prototype HPGe detector designed to be capable of achieving fine energy resolution and high event throughput at count rates in excess of one million counts per second. We report here on the concept, design, and initial performance of the first prototype device.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,