Article ID Journal Published Year Pages File Type
8173216 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2015 4 Pages PDF
Abstract
In this paper, we discuss results of avalanche photodiodes radiation tests for Projectile Spectator Detector at future Compressed Baryonic Matter experiment. The tests were carried out in Nuclear Physics Institute of ASCR in Řež using the cyclotron facility. Secondary neutron beam was used for irradiation because the main radiation damage in the Projectile Spectator Detector is caused by neutrons. Two types of the avalanche photodiodes from Zecotek and Ketek manufacturers were investigated. Special attention was given to the noise investigation and self-annealing after the irradiation. We have irradiated two Ketek PM3375 diodes with equivalent dose for 1 MeV neutrons equal to 2.5±0.2×1012 n/cm2, and single Zecotek MAPD-3N diode with equivalent dose for 1 MeV neutrons equal to 3.4±0.2×1012 n/cm2. All the types of the diodes have shown an increasing level of the noise after the irradiation. From that we can conclude that those avalanche photodiodes are not able to detect single photons anymore due to high noise levels.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , ,