Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8173277 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 7 Pages |
Abstract
In this study a procedure for range verification in proton therapy by means of a planar in-beam PET system is presented. The procedure consists of two steps: the measurement of the β+-activity induced in the irradiated body by the proton beam and the comparison of these distributions with simulations. The experimental data taking was performed at the CNAO center in Pavia, Italy, irradiating plastic phantoms. For two different cases we demonstrate how a real-time feedback of the delivered treatment plan can be obtained with in-beam PET imaging.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
A.C. Kraan, G. Battistoni, N. Belcari, N. Camarlinghi, M. Ciocca, A. Ferrari, S. Ferretti, A. Mairani, S. Molinelli, M. Pullia, P. Sala, G. Sportelli, A. Del Guerra, V. Rosso,