Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8173960 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 6 Pages |
Abstract
A reliable and simple-to-use algorithm was developed for the energy-calibration of double-sided silicon strip detectors (DSSSDs). It works by enforcing mutual consistency of p-side and n-side information for every detected event. The procedure does not rely on a dedicated data set for calibration and is robust enough to work fully automated without human supervision. The method was developed and applied to data from a DSSSD of the Lund-York-Cologne CAlorimeter (LYCCA) for the HISPEC experiment at FAIR. It has been tested on ions in the Aâ90 mass range at energies of Ekinâ300MeV/u.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
M. Reese, J. Gerl, P. Golubev, N. Pietralla,