8174489 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2015 11 Pages PDF
Abstract
Gamma-ray detector systems are important instruments in a broad range of science and new setup are continually developing. The most recent step in the evolution of detectors for nuclear spectroscopy is the construction of large arrays of detectors of different forms (for example, conical, pentagonal, hexagonal, etc.) and sizes, where the performance and the efficiency can be increased. In this work, a new direct numerical method (NAM), in an integral form and based on the efficiency transfer (ET) method, is used to calculate the full-energy peak efficiency of a single hexagonal NaI(Tl) detector. The algorithms and the calculations of the effective solid angle ratios for a point (isotropic irradiating) gamma-source situated coaxially at different distances from the detector front-end surface, taking into account the attenuation of the gamma-rays in the detector׳s material, end-cap and the other materials in-between the gamma-source and the detector, are considered as the core of this (ET) method. The calculated full-energy peak efficiency values by the (NAM) are found to be in a good agreement with the measured experimental data.