Article ID Journal Published Year Pages File Type
8174860 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2014 5 Pages PDF
Abstract
In this work, we investigated the ability of a high-purity germanium detector connected to a trapezoid-filter-based data acquisition system to reliably record signals in spite of high sample activities. By activating multiple Na2CO3 samples with different Na content, we were able to deduce efficiency, resolution and dead time of the system used as a function of the sample activity. Based on the results, we were able to find a setting which allows measurements of event rates up to 35 kHz per readout channel with an energy resolution of 0.3% at the 2754 keV 24Na line.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , , ,