Article ID Journal Published Year Pages File Type
8175644 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2014 5 Pages PDF
Abstract
Neutron measurements can be used to distinguish plutonium in metal or oxide form, a capability that is of great interest in nuclear nonproliferation, treaty verification, and other applications. This paper describes measurements performed on well-characterized samples of plutonium oxide and plutonium metal using the pulse-shape-discrimination-capable plastic scintillator EJ-299-33. Results are compared to those obtained with a same-sized detector cell using the liquid scintillator EJ-309. The same optimized, digital pulse shape discrimination technique is applied to both detectors and the neutron pulse height distributions are compared. Results show that the EJ-299-33 plastics can be successfully used for plutonium measurements, where the gamma ray to neutron detection ratio is much higher than for typical radioactive sources. Results also show that EJ-299-33 detectors can be used to characterize plutonium samples, specifically to discriminate between plutonium metal and oxide.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , ,