Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8175734 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2014 | 8 Pages |
Abstract
In many applications, such as the detection of ultra-high energy cosmic rays using the air fluorescence method, the number of photons incident on the detector must be known. This requires a precise knowledge of the absolute efficiency of the photodetectors used. We present an experimental setup for measuring the single photoelectron gain and efficiency of multi-anode photomultipliers with a total uncertainty on the order of a few percent. This precision is obtained by using a comparison to a NIST calibrated photodiode, and the presented method can be applied to both vacuum photomultiplier tubes and other photodetectors. This work is motivated by the need to calibrate the focal surface of the EUSO-Balloon instrument, which is a technical pathfinder for the future JEM-EUSO mission. A complete discussion of photomultiplier calibration is presented and the efficiency measurement technique is discussed in detail. A set of example results are shown in order to illustrate the key points of the method.