Article ID Journal Published Year Pages File Type
8176282 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2014 7 Pages PDF
Abstract
Liquid scintillator detector assemblies contain an inert nitrogen expansion volume to allow for expansion of the liquid with changing temperature. Measurements and Geant4 Monte Carlo simulations are performed to study the dependence of pulse height distribution shapes as a function of detector angle for two liquid scintillators assemblies filled with 97% organic-liquid cocktail and a 3% expansion volume. A 12.7-cm diameter by 12.7-cm long and a 7.6-cm diameter by 9.1-cm long EJ-309 liquid scintillator assemblies are investigated using a 137Cs gamma-ray source. Aside from the differences in dimensions, the detector assemblies also differed in the design of the active detector volume: there is no light guide in the 12.7-cm-diameter detector assembly, whereas the 7.6-cm-diameter detector contains a BK7 light guide between the scintillation liquid and optical coupling to the photomultiplier tube. Results for the 12.7-cm-diameter detector show a decrease in the position of the Compton edge ranges from 4% to 40% at detector orientations where the expansion volume exists between scintillating medium and the photomultiplier tube. Results for the 7.6-cm-diameter detector show that the position of the Compton edge is relatively unaffected at all detector orientations due to the presence of light guide.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , ,