Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8176317 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2014 | 6 Pages |
Abstract
A 4.36-mm-thick pixelated thallium bromide (TlBr) detector with Tl electrodes was fabricated from a crystal grown by the traveling molten zone method using zone-purified material. The detector had four 1Ã1Â mm2 pixelated anodes. The detector performance was characterized at room temperature. The mobility-lifetime products of electrons for each pixel of the TlBr detector were measured to be >2.8Ã10â3Â cm2/V. The four pixelated anodes of the detector exhibited energy resolutions of 1.5-1.8% full width at half maximum (FWHM) for 662-keV gamma rays for single-pixel events with the depth correction method. An energy resolution of 4.5% FWHM for 662-keV gamma rays was obtained from a reconstructed energy spectrum using two-pixel events from the two pixelated anodes on the detector.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Keitaro Hitomi, Toshiyuki Onodera, Seong-Yun Kim, Tadayoshi Shoji, Keizo Ishii,