Article ID Journal Published Year Pages File Type
8176412 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2014 6 Pages PDF
Abstract
Alpha particles are monitored for detecting nuclear fuel material (i.e., plutonium and uranium) at nuclear fuel facilities. Currently, for monitoring the airborne contamination of nuclear fuel, only energy information measured by Si-semiconductor detectors is used to distinguish nuclear fuel material from radon daughters. In some cases, however, such distinguishing is difficult when the radon concentration is high. In addition, a Si-semiconductor detector is generally sensitive to noise. In this study, we developed a new alpha-particle imaging system by combining a Si-PM array, which is insensitive to noise, with a Ce-doped Gd3Al2Ga3O12(GAGG) scintillator, and evaluated our developed system׳s fundamental performance. The scintillator was 0.1-mm thick, and the light guide was 3.0 mm thick. An 241Am source was used for all the measurements. We evaluated the spatial resolution by taking an image of a resolution chart. A 1.6 lp/mm slit was clearly resolved, and the spatial resolution was estimated to be less than 0.6-mm FWHM. The energy resolution was 13% FWHM. A slight distortion was observed in the image, and the uniformity near its center was within ±24%. We conclude that our developed alpha-particle imaging system is promising for plutonium detection at nuclear fuel facilities.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , ,