Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8179442 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2013 | 5 Pages |
Abstract
A transition radiation detector (TRD) prototype based on a single multiwire proportional chamber coupled with a small drift region was developed for the innermost part of the CBM-TRD subsystem. It preserves the same gas thickness for transition radiation absorption as the double-sided TRD prototype for which a pion misidentification probability of better than 1% for a six layer configuration was obtained. In the same time it fulfills the requirement of a high geometrical efficiency of CBM-TRD stations. The readout electrode geometry with triangular shaped pads gives access to a two-dimensional position information with a single TRD layer. The detector was tested with a mixed electron/pion beam of 1-5Â GeV/c momentum at the CERN PS accelerator. A pion misidentification probability of 1.18% for a six layer configuration based on this architecture was obtained. The two-dimensional position resolutions (along and across the pads) were measured. The pad signals were processed using a new front-end electronics called Fast Analog Signal Processor (FASP), designed for high-counting-rate environments. CADENCE simulations were used for further optimization of the FASP amplifier for operating this new architecture.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
M. PetriÅ, M. Petrovici, V. CÄtÄnescu, M. TârzilÄ, V. Simion, D. BartoÅ, I. Berceanu, A. Bercuci, G. Caragheorgheopol, F. Constantin, L. RÄdulescu, J. Adamczewski-Musch, S. Linev,