Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8179559 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2013 | 4 Pages |
Abstract
A position sensitive Cherenkov detector was built, consisting of 64 SiPMs with an active area of 3Ã3 mm2 and a pixel size of 100Ã100μm2. The sensitive area is increased by a light concentrator which consists of 64 pyramid-shaped funnels. These funnels have an entrance area of 7Ã7 mm2 and an exit area of 3Ã3 mm2, guaranteeing a sufficient position resolution e.g. for the barrel DIRC detector of the PANDA experiment at FAIR. The efficiency and uniformity of the light concentrator in combination with the SiPM array was tested by scanning the array in two dimensions, using a pulsed light beam. Results of these tests and comparison with simulations are given here.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Mariana Rihl, Stefan Enrico Brunner, Lukas Gruber, Johann Marton, Ken Suzuki,