Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8180284 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2013 | 4 Pages |
Abstract
A new method to determine the spatial resolution of a one-dimensional gaseous electron multiplication (GEM) prototype is demonstrated via an experimental test using an Fe-55 source with an adjustable slit. In this method, the total variance of the incident position Ïtot2, contributed by the GEM spatial resolution and the geometrical acceptance of the slit, is presented as a function of the event rate n under the full energy peak, with the latter being employed to calibrate the square of the slit width. Using a linear fit to the experimental data, a spatial resolution of Ïg=56±15μm(1Ï) is measured. Compared to the existing methods, the novel method presented here is robust considering the inaccuracy in the direct measurements of the slit widths, and hence greatly circumvents machining difficulties.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Rensheng Wang, Yan Huang, Zhigang Xiao, Zhao Zhang, Rong Wang, Haiyan Gao,