Article ID Journal Published Year Pages File Type
8181722 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2012 12 Pages PDF
Abstract
At the core of the AGILE scientific instrument, designed to operate on a satellite, there is the Gamma Ray Imaging Detector (GRID) consisting of a Silicon Tracker (ST), a Cesium Iodide Mini-Calorimeter and an Anti-Coincidence system of plastic scintillator bars. The ST needs an on-ground calibration with a γ-ray beam to validate the simulation used to calculate the energy response function and the effective area versus the energy and the direction of the γ rays. A tagged γ-ray beam line was designed at the Beam Test Facility (BTF) of the INFN Laboratori Nazionali of Frascati (LNF), based on an electron beam generating γ-rays through bremsstrahlung in a position-sensitive target. The γ-ray energy is deduced by difference with the post-bremsstrahlung electron energy [1], [2]. The electron energy is measured by a spectrometer consisting of a dipole magnet and an array of position sensitive silicon strip detectors, the Photon Tagging System (PTS). The use of the combined BTF-PTS system as tagged photon beam requires understanding the efficiency of γ-ray tagging, the probability of fake tagging, the energy resolution and the relation of the PTS hit position versus the γ-ray energy. This paper describes this study comparing data taken during the AGILE calibration occurred in 2005 with simulation.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , , , , , , , , , , ,