Article ID Journal Published Year Pages File Type
8208405 Applied Radiation and Isotopes 2018 11 Pages PDF
Abstract
The absorbed dose rate is the main characterization quantity for beta radiation. The extrapolation chamber is considered the primary standard instrument. To determine absorbed dose rates in beta radiation beams, it is necessary to establish several correction factors. In this work, the correction factors for the backscatter due to the collecting electrode and to the guard ring, and the correction factor for Bremsstrahlung in beta secondary standard radiation beams are presented. For this purpose, the Monte Carlo method was applied. The results obtained are considered acceptable, and they agree within the uncertainties. The differences between the backscatter factors determined by the Monte Carlo method and those of the ISO standard were 0.6%, 0.9% and 2.04% for 90Sr/90Y, 85Kr and 147Pm sources respectively. The differences between the Bremsstrahlung factors determined by the Monte Carlo method and those of the ISO were 0.25%, 0.6% and 1% for 90Sr/90Y, 85Kr and 147Pm sources respectively.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , ,