Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8208673 | Applied Radiation and Isotopes | 2018 | 5 Pages |
Abstract
The National Physical Laboratory Vinten 671 chamber was selected as a proving ground for a new radionuclide source model in the EGSnrc software. The computational Vinten model is validated against measurements of radionuclide artifacts whose activities were determined by absolute methods. The response of the Vinten chamber is first calculated as a function of gamma energy, but more strikingly, an explicit simulation of radionuclide decay was implemented and now permits the direct determination of a calibration factor, including additional effects due to all decay paths of the radionuclide. The Monte Carlo and experimental calibration factors are found to agree at the percent level, in absolute terms.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
R. Townson, F. Tessier, R. Galea,