Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8208722 | Applied Radiation and Isotopes | 2018 | 14 Pages |
Abstract
A 24 segmented HPGe coaxial detector was set up with a digitized data acquisition system (DAQ). The DAQ was composed of a digitizer (5 Ã 107sampling/s), a Field-Programmable Gate Array (FPGA), and a real time operating system. The Full Width Half Maximum (FWHM), rise time, signal characteristics, and spectra of a 137Cs source were evaluated. The data were processed using an in-house developed gamma-ray tracking system.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Jubong Han, K.B. Lee, Jong-Man Lee, S.H. Lee, Tae Soon Park, J.S. Oh,