Article ID Journal Published Year Pages File Type
8208916 Applied Radiation and Isotopes 2017 6 Pages PDF
Abstract
Experimental tests are presented that assess the cross-talk level among three scintillation detectors used as neutron counters exploiting the thermal neutron radiative capture on Cd. The measurements were done at the INES diffractometer operating at the ISIS spallation neutron source (Rutherford Appleton Laboratory, UK). These tests follow a preliminary set of measurements performed on the same instrument to study the effectiveness of this thermal neutron counting strategy in neutron diffraction measurements, typically performed on INES using squashed 3He filled gas tubes. The experimental data were collected in two different geometrical configurations of the detectors and compared to results of Monte Carlo simulations, performed using the MCNP code.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , ,