Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8209306 | Applied Radiation and Isotopes | 2016 | 4 Pages |
Abstract
A commercially available Fidelis ionization chamber was calibrated and assessed in PTB with activity standard solutions. The long-term stability and linearity of the system was checked. Energy-dependent efficiency curves for photons and beta particles were determined, using an iterative method in Excelâ¢, to enable calibration factors to be calculated for radionuclides which were not used in the calibration. Relative deviations between experimental and calculated radionuclide efficiencies are of the order of 1% for most photon emitters and below 5% for pure beta emitters. The system will enable TAEK-SANAEM to provide traceable activity measurements.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Emin Yeltepe, Karsten Kossert, Abdullah Dirican, Ole Nähle, Christiane NiedergesäÃ, Namik Kemal Åahin,