Article ID Journal Published Year Pages File Type
8209338 Applied Radiation and Isotopes 2016 4 Pages PDF
Abstract
In this work we present test procedures, approval criteria and results from two metrological inspections of a certified large volume 152Eu source (drum about 200 L) intended for calibration of HPGe gamma assay systems used for activity measurement of radioactive waste drums. The aim of the inspections was to prove the stability of the calibration source during its working life. The large volume source was designed and produced in 2007. It consists of 448 identical sealed radioactive sources (modules) apportioned in 32 transparent plastic tubes which were placed in a wooden matrix which filled the drum. During the inspections the modules were subjected to tests for verification of their certified characteristics. The results show a perfect compliance with the NIST basic guidelines for the properties of a radioactive certified reference material (CRM) and demonstrate the stability of the large volume CRM-drum after 7 years of operation.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, ,