Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8209373 | Applied Radiation and Isotopes | 2016 | 5 Pages |
Abstract
This paper reports on absolute activity measurements of iron-59 made at the National Metrology Institute of South Africa (NMISA) via 4Ï(LS)β-γ coincidence counting. The exercise formed part of an Asia Pacific Metrology Program (APMP) regional key comparison. Source data were analysed by the extrapolation technique for a number of gamma-ray window settings. In addition, a feasibility study was undertaken on a second technique; a non-extrapolation method based on a detection efficiency analysis. The reported activity concentration of the 59Fe solution was determined with a relative uncertainty of 0.28% (k=1), the uncertainty being due mainly to the rate vs. efficiency fitting process. The result from the non-extrapolation method was lower than that given by extrapolation by 0.33%, within two standard deviations. Possible reasons for the small discrepancy are discussed.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
M.W. van Rooy, M.J. van Staden, J. Lubbe, B.R.S. Simpson,