Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8209640 | Applied Radiation and Isotopes | 2015 | 7 Pages |
Abstract
Ra-223 is an alpha-emitter that is being used as a bone-seeking radiotherapeutic agent. The relatively large uncertainty on its evaluated half-life (0.26%, Bé et al., 2011) is an impediment to precision activity assays, which often involve measurements by various methods over time spans of days or weeks. We have performed two series of measurements using an ionization chamber (IC) and a NaI(Tl) well counter (γ-wc) to determine new, precise values for the 223Ra half-life. We have endeavored to realistically assess the uncertainties on the derived half-lives, looking beyond the fit uncertainties to identify uncertainty components acting on multiple timescales. We recovered respective values of 11.447(6) d and 11.445(13) d from the IC and γ-wc measurements. Our values are in accord with the evaluated value of 11.43(3) d, but with smaller combined uncertainties.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Denis E. Bergeron, Ryan Fitzgerald,