Abstract
The MCNPX code was used to determine the efficiency of a N-type HPGe detector after two decades of operation. Accounting for the roundedness of the crystal`s front edges and an inhomogeneous description of the detector's dead layers were shown to achieve better agreement between measurements and simulation efficiency determination. The calculations were experimentally verified using point sources in the energy range from 50Â keV to 1400Â keV, and an overall uncertainty less than 2% was achieved. In order to use the detector for different matrices and geometries in radioactivity, the suggested model was validated by changing the counting geometry and by using multi-gamma disc sources. The introduced simulation approach permitted the revaluation of the performance of an HPGe detector in comparison of its initial condition, which is a useful tool for precise determination of the thickness of the inhomogeneous dead layer.