Abstract
This paper compares trace element concentrations (Ca, K, Sr, Fe, Mn, Zn, Ni, Cu, Co and Cr) in 27 Sudanese medical plants determined in parallel by PIXE and ICP-OES to get information on which technique is preferable at different matrices and element concentrations. PIXE correlates well to ICP-OES for Sr, Mn, Ca, K, Zn and Fe determinations. ICP-OES seems to be the superior technique over PIXE when measuring low concentrated elements (chromium, cobalt, nickel and copper) in the medicinal plants.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Ammar Mubark Ebrahim, M.A. Etayeb, Khalid H., Manale Noun, M. Roumie, B. Michalke,