Abstract
The predicted biological dose delivered to a tumor region, based on the variable RBE inferred from the data, varies significantly with respect to the clinically used constant RBE of 1.1. The significant RBE increase at the distal end suggests also a potential to enhance optimization of treatment modalities such as LET painting of hypoxic tumors. The study highlights the limitation of adoption of a constant RBE for proton therapy and suggests approaches for fast implementation of RBE models in treatment planning.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Pankaj PhD, Thomas I. MSci, Francesca M. MSc, Lorenzo PhD, Frederick J. PhD, Fiona MSci, Stephen J. PhD, Joy N. PhD, Giuseppe Antonio Pablo PhD, Francesco PhD, Kevin M. PhD, Giuseppe PhD,