Abstract
The requirements of quality assurance (QA) for both brachytherapy and imaging devices are well-defined, but image-guided brachytherapy has raised new issues. Image guidance in brachytherapy involves the transition from reference point dosimetry using films to volumetric imaging such as computed tomography, ultrasonography, and magnetic resonance imaging for treatment planning and guidance of applicator, needle, or seed placement. The QA of these devices might not reflect the conditions of use in brachytherapy or the requirements of brachytherapy treatment planning. Image interpretation becomes much more important with image-guided brachytherapy. The success of a procedure could depend on the interpretation of a single image in a calibration phase done under the time pressures of the operative setting. This change has implications at the level of treatment, the process, and the field of brachytherapy as a whole. The QA concerns arising from brachytherapy procedures using ultrasound, computed tomography, and magnetic resonance imaging guidance are discussed, as are the problems associated with using imaging in an interventional setting. This report was intended to indicate the QA concerns arising from the convergence of brachytherapy and imaging-highlighting areas in which technical improvements are needed.