Abstract
The dose measured with MOSFETs and GAFROMIC EBT2 films agreed well with the expected dose. For both methods, the dose to the thoracic wall, lungs and heart for left sided patents was lower than 2.5â¯Gy even when 12â¯MeV was applied. The positioning time of GAFCHROMIC films is negligible and based on our results we recommend its use as a standard tool for patient quality assurance during breast cancer IORT.