Article ID Journal Published Year Pages File Type
8249880 Radiation Measurements 2018 11 Pages PDF
Abstract
TL glow curves recorded with heating rates in the range 0.02-5 K s−1 showed peak positions varying up to 60 °C between systems at the highest heating rates, attributed to temperature calibration errors and/or thermal lag. Kinetic parameters derived from the complete data set show a large spread, covering the ranges ∼0.5-1.2 eV and 106-1017 s−1 for E and s. In most cases, interlaboratory variations exceeded those of replicate measurements within individual laboratories. Signal lifetimes at 20 °C derived from the isothermal decay (∼59 min) and initial rise methods (at low heating rates; ∼60-80 min) most closely match the value directly measured at 20 °C from within two luminescence readers (∼70 min). Finally, we discuss the consequences of these findings for dosimetry and dating using luminescence signals and possible ways to reduce systematic errors in laboratory measurements of kinetic parameters.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation