Abstract
Diamond based microdosimeter shows a well-defined active volume. Homogeneity of the response was estimated at about 7% and linked to structural defects of the diamond surface as deduced by AFM inspection. The detector response shows a good linear behavior for ions of different kind and energy, indicating that the detector is suitable for measuring a wide range of particles and LET i.e. 100â¯Ã·â¯3000â¯keV/μm. Finally, microdosimetric capabilities of the diamond based microdosimeter were preliminarily tested in low LET radiation fields (i.e. protons beam).
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
C. Verona, G. Magrin, P. Solevi, M. Bandorf, M. Marinelli, M. Stock, G. Verona Rinati,