Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8253137 | Zeitschrift für Medizinische Physik | 2018 | 12 Pages |
Abstract
The second aim of this study has been a novel, semi-empirical approach to calculate the field-size dependent output correction factors of small photon detectors by convolving film measured true dose profile data with measured lateral response functions of the detectors. To achieve this, the set of previously published 2D lateral dose response functions was complemented by those of the novel detectors PTW PinPoint chamber 31022 (PTW Freiburg, Freiburg, Germany), Razor chamber and Razor Diode (IBA Dosimetry, Schwarzenbruck, Germany). The output correction factors calculated from the lateral dose response functions closely fit with the directly measured output correction factors, thus supporting the latter by an independent method.
Related Topics
Physical Sciences and Engineering
Engineering
Biomedical Engineering
Authors
Daniela Poppinga, Björn Delfs, Jutta Meyners, Dietrich Harder, Björn Poppe, Hui Khee Looe,