Article ID Journal Published Year Pages File Type
848702 Optik - International Journal for Light and Electron Optics 2015 4 Pages PDF
Abstract
An optical spectrometer based on the Fourier transform is presented. Both the measured light and reference light are illuminated into a Michelson interferometer with angle difference between two mirrors, and the Fourier transform spectra of the interference fringes are obtained by using charge-coupled device (CCD) and the field-programmable gate array (FPGA). The peak point in the Fourier transform spectra of the reference light is used to calibrate the measured light, and the wavelength of the measured light can be detected by calculating the Fourier transform spectra between the measured light and the reference light. Due to the Michelson interferometer with angle difference between two mirrors, the Fourier transform spectrometer is realized without any mechanical movement, and the accuracy of the optical spectrometer is improved by propagating two lights into the common optical path in the same Michelson interferometer. The experiment shows that the maximum error is 1.15 nm. The proposed method possesses high performance, high reliability, and low cost.
Related Topics
Physical Sciences and Engineering Engineering Engineering (General)
Authors
, , ,