Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
862668 | Procedia Engineering | 2011 | 5 Pages |
Abstract
In the traditional human bioelectrical impedance measurement methods, measurement approach is simplex and accuracy is low. A multi-frequency and multi-segment human bioelectrical impedance measurement method is proposed, which adopts eight electrodes segments measurement technique and direct frequency synthesis method to produce multi-frequency excitation signal, and then achieves multi-frequency and multi-segment measurement of bioelectrical impedance. The results of experiments show that this method effectively improves the measurement accuracy; the relative error of measured impedance magnitude is less than 1% while the absolute error of phase is less than 1 degree.
Related Topics
Physical Sciences and Engineering
Engineering
Engineering (General)