A 71 year old male who had undergone extra-anatomic bypass grafting between the ascending aorta and the thoraco-abdominal aorta at 41 years of age for aortic coarctation was admitted with back pain and dyspnea. A 16 mm Cooley double velour knitted polyethylene terephthalate (PET) graft was used in the initial operation in 1983. Computed tomography showed disruption of the initial PET graft perforating the right atrium, and a pseudoaneurysm at the distal anastomosis. The patient was in acute cardiac failure because of left to right shunting. A two stage operation was performed. The first stage comprised emergency re-grafting and right atrium repair, and the second stage re-grafting for the pseudoaneurysm. The patient is doing well 48 months post-operatively; however, monitoring of the patient will continue for potential PET graft rupture.
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine
Authors
Akihito Matsushita, Yu Tsunoda, Takashi Hattori, Wahei Mihara,