Article ID Journal Published Year Pages File Type
867740 Biosensors and Bioelectronics 2011 5 Pages PDF
Abstract

An amperometric glucose biosensor based on an n-alkylamine-stabilized palladium nanoparticles (PdNPs)-glucose oxidase (GOx) modified glassy carbon (GC) electrode has been successfully fabricated. PdNPs were initially synthesized by a biphase mixture of water and toluene method using n-alkylamines (dodecylamine, C12-NH2 and octadecylamine, C18-NH2) as stabilizing ligands. The performance of the PdNPs-GOx/GC biosensor was studied by cyclic voltammetry. The optimum working potential for amperometric measurement of glucose in pH 7.0 phosphate buffer solution is −0.02 V (vs. Ag/AgCl). The analytical performance of the biosensor prepared from C18-PdNPs-GOx is better than that of C12-PdNPs-GOx. The C18-PdNPs-GOx/GC biosensor exhibits a fast response time of ca. 3 s, a detection limit of 3.0 μM (S/N = 3) and a linear range of 3.0 μM–8.0 mM. The linear dependence of current density with glucose concentration is 70.8 μA cm−2 mM−1. The biosensor shows good stability, repeatability and reproducibility. It has been successfully applied to determine the glucose content in human blood serum samples.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,