Article ID Journal Published Year Pages File Type
867867 Biosensors and Bioelectronics 2010 6 Pages PDF
Abstract

We developed an electrochemical-sensing device for continuous monitoring extracellular hydrogen peroxide (H2O2). The device consists of an indium-tin-oxide electrode coated with osmium-polyvinylpyridine gel polymer containing horseradish peroxidase (Os-HRP) and a poly-dimethyl siloxane well to house the cells on the chip. Granulocyte-like differentiated HL-60 cells were accommodated in the well and stimulated with phorbol 12-myristate 13-acetate (PMA), which triggered the generation of H2O2. The extracellular H2O2 released from the cells was enzymatically reduced at the Os-HRP-modified electrode chip using Os(II) as an electron donor, resulting in reduction current responses by the device. The reduction current increased immediately upon PMA stimulation and this current transient was similar to that obtained by conventional chemiluminescence assays using sodium luminol. Apocynin, an inhibitor of NADPH oxidase activation, eliminated both the electrochemical and chemiluminescence signals. On the other hand, superoxide dismutase (SOD) increased the amperometric signals and catalase (CAT) decreased, whereas SOD decreased luminescence emission and CAT did not. These results were in accordance with the expected reaction mechanism, and strongly indicate that this new electrochemical-sensing device successfully detects extracellular H2O2 production.