Article ID Journal Published Year Pages File Type
868423 Biosensors and Bioelectronics 2010 5 Pages PDF
Abstract

Novel electrochemical platform based on Pt nanoparticle modified ordered three-dimensional gold nanowire arrays (PtNP/NAEs) for the amperometric sensing of H2O2 and glutamate is developed. Pt nanoparticle (PtNP) is fabricated by electrodeposition onto the 3D nanowires and characterised using scanning electron microscopy (SEM) and cyclic voltammetry. The deposited nanoparticles have an average size of 20 nm. The PtNP/NAE shows a linear response of up to 20 mM for H2O2 detection with a sensitivity of 194.60 μA mM−1 cm−2 at 20 °C. It can detect 1 μM (S/N = 3) of H2O2 at normal condition without using any enzyme or mediator. Analytical performance of this electrode is tested by immobilising glutamate oxidase (GlutOx) through cross-linking in the matrix of bovine serum albumin (BSA), Nafion and glutaraldehyde. At physiological pH, the biosensor showed the sensitivity of 10.76 μA mM−1 cm−2, with a linear range of up to 0.8 mM.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,