Article ID Journal Published Year Pages File Type
868820 Biosensors and Bioelectronics 2010 5 Pages PDF
Abstract

A high-precision optical biosensor technique capable of independently determining the refractive index (RI) of liquids is presented. Photonic crystal surface waves were used to detect surface binding events, while an independent registration of the critical angle was used for accurate determination of the liquid RI. This technique was tested using binding of biotin molecules to a streptavidin monolayer at low and high biotin concentrations. The attained baseline noise is 5×10−135×10−13 m/Hz1/2 for adlayer thickness changes and 9×10−89×10−8 RIU/Hz1/2 for RI changes.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,