Article ID Journal Published Year Pages File Type
872806 Journal of Biomechanics 2009 7 Pages PDF
Abstract

A new definition of the experimental wear factor was established and reported as a function of cross-shear motion and contact pressure using a multi-directional pin-on-plate wear testing machine for conventional polyethylene in the present study. An independent computational wear model was developed by incorporating the cross-shear motion and contact pressure-dependent wear factor into the Archard's law, in additional to load and sliding distance. The computational prediction of wear volume was directly compared with a simulator testing of a polyethylene hip joint with a 28 mm diameter. The effect of increasing the femoral head size was subsequently considered and was shown to increase wear, as a result of increased sliding distance and reduced contact pressure.

Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering