Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8737316 | Diagnostic Microbiology and Infectious Disease | 2018 | 20 Pages |
Abstract
The Loopampâ¢MTBC detection kit (TB-LAMP) was designed to replace the sputum smear microscopy for the diagnosis of pulmonary tuberculosis. We evaluated its performance at a peripheral laboratory in Vietnam. The sensitivity of TB-LAMP was 45.5% (28.1%-63.6%), which was equal to 3-sputum smear microscopy but lower than that of Xpert MTB/RIF (87.8% [71.8%-96.6%]). In patients with culture-confirmed TB, sensitivity was 80% (51.9%-95.7%) in smear-positive and 16.7% (3.5%-41.4%) in smear-negative cases. The specificity of TB-LAMP was 95.1% (92.7%-96.9%), which was lower than that of smear microscopy (98.9% [97.5%-99.7%]) and Xpert MTB/RIF (99.3% [98.1%-99.9%]) (P < 0.05). The probability of TB detection by TB-LAMP was more influenced by sample quality and viscosity than were smear microscopy, Xpert MTB/RIF, and culture. The present data do not support the use of TB-LAMP as a replacement test for smear microscopy in peripheral laboratories.
Keywords
Related Topics
Life Sciences
Immunology and Microbiology
Applied Microbiology and Biotechnology
Authors
Van Anh Thi Nguyen, Hung Van Nguyen, Thuong Van Dinh, Ha Hoang Du, Chinh Nhu Do, Guy B. Marks, Nhung Viet Nguyen,