876580 Medical Engineering & Physics 2011 10 Pages PDF
Abstract

There are methodological concerns with combined use of transcranial Doppler (TCD) and Finapres to measure dynamic cerebral autoregulation. The Finapres calibration mechanism (“physiocal”) causes interruptions to blood pressure recordings. Also, TCD is subject to signal loss due to probe movement. We assessed the effects of “physiocals” and TCD signal loss on transfer function estimates in recordings of 45 healthy subjects. We added artificial “physiocals” and removed sections of TCD signal from 5 min Finapres and TCD recordings. We also compared transfer function results from 5 min time series with time series as short as 1 min. Accurate transfer function estimates can be achieved in the 0.03–0.07 Hz band using beat-by-beat data with linear interpolation, while data loss is less than 10 s. At frequencies between 0.07 and 0.5 Hz, transfer function estimates become unreliable with 5 s of data loss every 50 s. 2 s data loss only affects frequency bands above 0.15 Hz. Finally, accurate transfer function assessment of autoregulatory function can be achieved from time series as short as 1 min, although gain and coherence tend to be overestimated at higher frequencies.

Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering