Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8799582 | The Journal of Arthroplasty | 2018 | 28 Pages |
Abstract
The results of this study suggest that mixing components from different manufacturers may affect the taper connection strength and could reduce the stability. As safety should be a high priority in patient treatment, any potential risks should be avoided. Therefore, mixing and matching of heads and femoral stems from different manufacturers cannot be recommended.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Ulrike MSc, Benjamin MD, Steffen MEng, Robert MSc, J. Philippe PhD,