Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8817480 | Biomedical and Environmental Sciences | 2018 | 4 Pages |
Abstract
To examine accuracy of body composition predicted by the Multi-Frequency Bioelectric Impedance Analysis (MF-BIA) compared with the Dual-energy X-ray Absorptiometry (DXA) in adults with obese. We measured body composition of 749 adults with obese both by the MF-BIA and DXA. The Lin's concordance correlation and the Bland-Altman plots were used to examine the consistency. The concordance correlation coefficient of %BF between the MF-BIA and DXA in men and women was 0.560, and 0.669, respectively. Compared with the DXA, the MF-BIA significantly underestimated %BF by 4.33% in men (P < 0.001), however overestimated %BF by 0.50% in women (P < 0.001). After corrected by the correction equations established in this study, the differences were significantly decreased. Therefore, the MF-BIA (TANITA MC-180) may need to be corrected in estimating body composition for adults with obese.
Related Topics
Health Sciences
Medicine and Dentistry
Public Health and Health Policy
Authors
WANG Zheng He, YANG Zhong Ping, WANG Xi Jie, DONG Yan Hui, MA Jun,