Article ID Journal Published Year Pages File Type
8821759 Clinical Imaging 2017 7 Pages PDF
Abstract
This study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64 ± 0.15 μGy and 49.11 ± 1.46 μGy, respectively (p < 0.001). The mean SNR values for mini-DI and DR were 233.2 ± 5.1 and 31.6 ± 1.2, and 10% MTF values were 131 and 161 μm. A newly developed mini-DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care.