Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8822583 | European Journal of Radiology | 2018 | 7 Pages |
Abstract
Statistically significant equivalence was shown with an equivalence margin of 0.12% points for cancer detection rate and with an equivalence margin of 0.40% points for specificity. A marked patient dose savings in grid-less acquisition of up to 36% compared to grid-based acquisition was achieved. It can be concluded that grid-less acquisition with software scatter correction is an alternative to grid-based acquisition in mammography.
Related Topics
Health Sciences
Medicine and Dentistry
Radiology and Imaging
Authors
Ahmed Jibril Abdi, Andreas Fieselmann, Heiderose Pfaff, Thomas Mertelmeier, Lisbet Brønsro Larsen,