Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8933624 | British Journal of Anaesthesia | 2013 | 6 Pages |
Abstract
Arterial pressure measurement based on radial artery tonometry using the TL-200pro technology is feasible in medical ICU patients. The TL-200pro system is capable of providing MAP values with high accuracy (low mean difference) and precision (narrow limits of agreement) compared with MAP measured invasively using a femoral arterial catheter. The TL-200pro technology is promising for the measurement of SAP and DAP but further development is necessary to improve accuracy and precision.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Anesthesiology and Pain Medicine
Authors
B Saugel, A.S. Meidert, A Hapfelmeier, F Eyer, R.M. Schmid, W Huber,