Article ID Journal Published Year Pages File Type
8939322 British Journal of Anaesthesia 2006 10 Pages PDF
Abstract
The version of the A-Line™ monitor used in this study does not exclude contaminated MLAEP signals. Previous publications involving this version of the A-Line™ monitor (as opposed to the newer A-Line/2™ monitor series) should be reassessed in the light of these findings. Before exclusively MLAEP-based monitors can be evaluated as suitable monitors of depth of anaesthesia, it is essential to ensure that inbuilt validity tests eliminate contaminated MLAEP signals.
Related Topics
Health Sciences Medicine and Dentistry Anesthesiology and Pain Medicine
Authors
, , , , , , ,